Altrix™ - Preventive Maintenance

Modified on Wed, 01 Apr 2020 at 12:21 PM

Q: Which items should be inspected and replaced during the preventive maintenace?


A: Remove the product from service before you perform the preventive maintenance inspection. Check all items listed during

annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks

more often based on your level of product usage. Service only by qualified service technician.


Note: Complete functional checks as indicated in the maintenance manual.


Inspect all of the following items:


_____All fasteners secure

_____Power cord and plug for fraying

_____ Condition of covers and push handle for damage

_____ Hose ports are operational

_____ Ground chain attached

_____ LCD is not cracked

_____ LCD operational

_____ Touch screen operational

_____ Wheels for smooth operation

_____ Rear caster wheels for free swivel action

_____ Both rear wheels lock secure when applied

_____ Front and rear wheels are not loose or wobbly

_____ Battery backup, functional check

_____ Alarm system - visual and audible, functional check

_____ High temp cutout checks

_____ Water temperature and flow verification, functional check

_____ Probe resistance, functional check

_____ Clear RFU codes

_____ Ground impedance not more than 100mΩ (millohms)

_____ Current leakage not more than 300 (microamps)

_____ Integrity of all clamps and clamped joints located in the air elimination circuit.


Replace the following on an annual basis:


_____ 9 V battery

_____ Condenser inlet filter

_____ Air eliminator hose

_____ Inlet filter

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select atleast one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article