LIFEPAK® 20/20e - Maintenance Prompt Interval

Modified on Mon, 30 Mar, 2020 at 4:46 PM

Q: How to set/reset maintenance prompt interval?


A: Chec below procedures:


Setting a Maintenance Prompt Interval

The LIFEPAK 20 defibrillator/monitor can be set to display a screen message that alerts the user when the maintenance prompt interval date has passed. The screen message MAINTENANCE DUE appears on the screen for the first 10 minutes after the device is powered on. The device maintenance interval can be turned off or set to 3 months, 6 months, or 12 months; the factory default is OFF, but it can be activated by a service technician.


To change the scheduled maintenance interval:

  • Display the SERVICE menu
  • Select MAINT PROMPT. The SERVICE/MAINT PROMPT submenu appears showing the current prompt date for scheduled maintenance (if set)
  • Select INTERVAL. The interval choices are: OFF, 3 MONTHS, 6 MONTHS, and 12 MONTHS
  • Select the desired interval
  • Turn the device OFF

Resetting the Maintenance Prompt

After completing scheduled maintenance, reset the maintenance prompt counter to clear the MAINTENANCE DUE message and begin the count for the next scheduled maintenance.


To turn off or reset the scheduled maintenance prompt:


  • Display the SERVICE menu
  • Select MAINT PROMPT. The SERVICE/MAINT PROMPT menu appears, showing the current prompt date for scheduled maintenance
  • Select RESET. The prompt date is revised to the next scheduled maintenance date
  • Turn the device OFF


Errors and omission excepted

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article